Description
1. Assess Micro Soot Sensor General Health/Functionality
- Temperatures, Pressures, Flows
- Microphone & Laser Linearity
- Span Check
- Window Pollution & Zero Signal Checks
- Internal Tubing Assessment
2. Annual Instrument Preventative Maintenance & Calibration
- Calibration of Pressure Sensors, Flow Meters, and Thermocouples
- Calibration of Analog Outputs
- Clean gas cell and windows
- Replace measuring unit internal tubing if necessary
- Replace sample line tubing
- Conditioning Unit Pump
- Customer Provides
- AVL PM Kit
- Calibrated Reference Pressure Sensors & Thermal Well
3. Troubleshoot and Repair Instrument failures
4. Perform Instrument Calibration Check
*Requires Customer provided Soot Source
Show More